I. Function Introduction

Online Hydrogen-in-Oxygen Analyzer KX-5100SO2 is a trace oxygen detection instrument using imported electrochemical sensors, featuring high stability, reproducibility, and accuracy.
Hydrogen-cooled units in power plants effectively reduce the thermal loss of generator windings through hydrogen cooling, thereby improving power generation efficiency. Losses caused by winding impedance are referred to as ventilation losses. Even generators slightly contaminated by oil, water, and air will experience increased ventilation losses. Data shows that a 1% drop in hydrogen purity results in a 12% increase in loss, so improving hydrogen purity means reducing production costs. A decrease in hydrogen purity and an increase in moisture content can also adversely affect the safe operation of generators.
The measurement of hydrogen content in mixed gases in petrochemical production is highly significant for processes such as cracking and hydrogenation. The oxygen content in hydrogen directly affects key performance indicators of the product. Additionally, measuring hydrogen purity is essential in the synthetic ammonia industry.
Recommended applications for the Online Hydrogen-in-Oxygen Analyzer KX-5100SO2: Hydrogen-cooled units in power plants, hydrogen production stations, petroleum processing, synthetic ammonia industry, metallurgical industry, electronics industry.

II. Main Features

Uses imported electrochemical sensors

LCD display, Chinese menu, easy maintenance

Provides multiple signal outputs

Users can perform on-site calibration to ensure reliable use

Supports gas pretreatment function

Gas pressure stabilization, flow regulation indication

 

 

III. Technical Specifications

Sensor

Original imported electrochemical sensor

Range

0~1%

Typical Accuracy

±0.01%

Reproducibility

±0.02%

Sample Gas Flow

100~500mL/min

Response Speed

10~15 seconds

Operating Temperature

-20℃~50℃

Operating Pressure

0.1~1.0Mpa

Signal Output

4~20mA(0~21% or custom), ±0.05mA 300Ω

Display

LCD remote display

Alarm Output

1 high alarm open contact output

Power Supply

220VAC±10%  50Hz  Power ≤20W

Accessories

Zener safety barrier

Inlet/Outlet Connection

φ6, φ8, φ10 adapter

Display Unit Dimensions

External dimensions 144×108×135  Cutout 139×101

Field Measurement Unit Dimensions

300×380×70

 

 

Note: Online Hydrogen-in-Oxygen Analyzer KX-5100SO2 can be customized according to user requirements. For details, please contact Zhengzhou Kaixuan Technology Co., Ltd.

Industry Category Measurement-Analysis-Instruments
Product Category
Brand:
Spec:
Stock:
Origin: China / Henan / Zhengzhoushi
About Toocle.com - Partner Programme - Old Version
Copyright © Toocle.com. All Rights Reserved.
(浙)-经营性-2023-0192